Automatic Proximate Analyzer (TGA)

Thermal Gravimetric Analyzer

Electronic Balance Mettler Toledo (USA) with precision 0.0001g

At the same test :

  • Ash
  • Inner Moisture
  • Volatile Matter
  • Net Calories
  • Fixed Carbon

 

Application Area

It is applicable in the industry such as electrical power, coal, commerce inspection, environment protection, metallurgy, papermaking, petrochemical, earth probe and the research institute or quality inspecting organization to test the content of inner moisture, ash, LOI, and volatile matter in the coal and other materials. With the value of Mar (measured by moisture tester), the instrument can also automatically calculate Calorific Value and the content of fixed carbon and hydrogen in the samples.

Technical Parameters

Furnace temperature: 100 °C ~1000 °C
Temperature controlling precision: ± 5 °C
Weight of sample: 0.5g ~ 0.7g
Sample number: 19 samples per time + 1 reference sample
Testing time: about 20 minutes per sample
Power Supply: 220V, 50/60 Hz
Power: ≤ 2 KW
Dimension: 400 mm x 510 mm x 640 mm
Weight: 65kg

Features

  1. Integrated with high precision electronic scale. Using thermogravimetry. Quickly, automatically and continuously tests 11 samples in every experiment. The air-dry basis moisture, ash and volatile constituent can be tested out in 20 minutes. Automatically tests the Calorific Value, fixed carbon and hydrogen content of the sample.
  2. Adopting the unique integrated configuration of “three temperature grade, double furnace”, the instrument can work in parallel, and hence to reduce the testing time and the constant temperature area in the furnaces do not interfere with each other. So the testing accuracy of air-dry basis moisture, ash, and volatile constituent can be ensured.
  3. Using the advanced heating element and heat preservation materials to ensure the furnace body in stable condition. Meanwhile, the temperature rise more quickly, temperature control will be more precise, and malfunction rate lower.
  4. With the advanced and appropriate dust suction and fume exhaust devices, the testing results will be more precise and the laboratory cleaner.
  5. Data collection, calculation, print, and restoration are automatically done by computer.
  6. Adopting precise reference source, low temperature drift and high precision operational amplifier, and 12-Bit A/D conversion chip in the controlling card, the instrument has the functions of cold end compensation, over-temperature protection both in hardware and software.
  7. Adopting the advanced USB technology, one computer can control several instruments. And the synthesized instrument can be composed combined with other instruments.
  8. Completely support Windows systems; high stability; can link electronic scale and share data in long distance through network.